CLT10sc Total Chlorine Analyzer (Panel Only) with pHD Differential Sensor

CLT10sc Total Chlorine Analyzer (Panel Only) with pHD Differential Sensor
Product #: LXV45B.99.13022
Unit Price Contact Hach
Ships within 1 week

EPA-approved Free or Total Chlorine Analysis

The CLF10 sc and CLT10 sc chlorine analyzers are compliant with EPA Method 334.0, ensuring that you not only have precise process control, but reportable results as well.

No Reagents to Replace, No Waste Stream

Hach’s amperometric chlorine technology does not use reagents, eliminating the need for a sanitary drain and saving you the cost of monthly reagent replacement.

Chlorine Analysis Made Easy

The CLF10 sc and CLT10 sc are plug-and-play with Hach’s digital controllers, ensuring easy setup. Available with exclusive self-diagnostics, these instruments will alert the user to changes in the process or upcoming maintenance. Hach’s Cal Watch feature will provide a warning for pH or chlorine calibration changes, and a non-contacting flow sensor will notify users of insufficient sample flow.

What's included?

Hach CL10sc Panel, 1 M digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.

Specifications

Accuracy: Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2: Free Chlorine: ± of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3: Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4: Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection: 5 pin, M12 connector
Cable Length: 1 m (between gateway and sc controller)
Calibration Method: 1-point or 2-point (zero and slope) calibration
Certifications: CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Connection Drain Line: 1/2 in ID
Connections: 1/4 in OD
Controller: Panel only
Controller Compatibility: sc controller platform
Description : CLT10sc Total Chlorine Analyzer (Panel Only) with pHD Differential Sensor
Dimensions : mm x 25 mm (7.68 in x 0.98 in)
  Sensor Only: 195
Fitting type: USA
Includes: CL10sc Panel, 1 m digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.
Interferences: Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Length: Sensor: 195 mm
Lower Limit of Detection (LOD): 30 ppb (0.03 ppm)
Material: Corrosion-resistant materials, fully-submersible
(stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement method: Reagentless, electrochemical, three-electrode amperometric system
Mounting: Panel or Wall Mounting
Operating Temperature Range: 0 - 45 °C
Options: pHD Differential Sensor
Parameter: Total Chlorine
pH Monitoring Required?: with pHD Differentinal
pH Range: 4 - 9
Power requirements (Voltage): None
Pressure Range: 0.5 bar, no pressure impulses and/or vibrations
Range: 0 - 20 ppm Cl₂
Repeatability: 30 ppb or 3%, whichever is greater
Response Time: Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
  Total Chlorine: 100 s or less for 90% change (T90)
at a stable temperature and pH
Sample Flow Rate: 30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature: 5 - 45°C (41 - 113°F)
Storage conditions: -20 °C to 60 °C dry
Warranty: 12 months
Weight: 11.18 kg Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?: Hach CL10sc Panel, 1 M digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.