Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / UL certified
Cycle Time: 10 min (dilution + 5 min.) per parameter
Digital outputs: Modbus RS485
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 64 mm
Exceedance Tracking: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Calmagite/EDTA method: Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA.
HNB method: dissolved Copper Cu(II) > 2 mg/L, Iron Fe(II) > 20 mg/L, Manganese > 10 mg/L, Zinc > 5 mg/L, Lead > 5 mg/L, Aluminium > 5 mg/L and Tin Sn(IV) interfere. Orthophosphate precipitates Calcium at the pH of the test. Strontium and Barium give a positive interference and alkalinity in excess of 300 mg/L may cause an indistinct end point in hard waters.
Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD): ≤ 2.5 µg/L CaCO3
Material: Hinged part: Thermoform ABS, door: plexiglass
Wall section: galvanised steel, powder coated
Measurement method: Total Hardness: colorimetric measurement at 610 nm using calmagite/EDTA
Calcium Hardness: colorimetric measurement at 610 nm using hydroxynaphthol blue/EDTA
Number of sample streams: 1 stream
Output: Modbus RS485
Optional: max. 8 active 4 - 20 mA max. 500 Ohm load outputs, RS232, Modbus TCP/IP
Parameter: Total Hardness, Calcium, Magnesium
Power: 110 - 240 VAC, 4 A, 50/60 Hz
Max. power consumption: 150 VA
Precision: Better than 2% full scale range for standard test solutions
Protection Class: Analyser cabinet: IP55 / Panel PC: IP66
Range: TH/Ca/Mg: 0 - 1 mg/L CaCO3

Optional:
0 - 0.1 mg/L
0 - 0.25 mg/L
0 - 0.5 mg/L
0 - 4 mg/L (with internal dilution)
0 - 8 mg/L (with internal dilution)
0 - 100 mg/L (with internal dilution)
Reagent Requirements: Keep between 10 - 30 °C
Sample Flow Rate: 100 - 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 1 year
Weight: 25 kg