EZ6301 Total Cadmium + Total Lead Analyser, 1 stream, Modbus RS485

EZ6301 Total Cadmium + Total Lead Analyser, 1 stream, Modbus RS485
Product #: EZ6301.990B1C02
CAD Price: Contact Hach

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / UL certified
Cooling water: Flow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
Cycle Time: 20 minutes (dilution + 5 min.)
Demineralised water: For rinsing and/or dilution
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 64 mm
Exceedance Tracking: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Thallium (III), tin (VI), organic matter may interfere. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD): ≤ 1 µg/L
Material: Hinged part: Thermoform ABS, door: plexiglass
Wall section: Galvanised steel, powder coated
Measurement method: Stripping voltammetry using carbon electrode
Number of sample streams: 1 stream

Optional:
1 to 6 streams
Output: Modbus RS485

Optional:
Active 4 - 20 mA max. 500 Ohm load, 1 to 8 outputs
RS232, Modbus TCP/IP
Parameter: Cadmium, total; Lead, total
Power: 110 VAC / 60 Hz
Precision: Better than 5% full scale range for standard test solutions
Protection Class: Analyser cabinet: IP55 / Panel PC: IP65
Range: 0 - 100 µg/L Cd, 0 - 100 µg/L Pb

Optional:
0 - 400 µg/L Cd, 0 - 400 µg/L Pb (with internal dilution)
0 - 1000 µg/L Cd, 0 - 1000 µg/L Pb (with internal dilution)
0 - 2000 µg/L Cd, 0 - 2000 µg/L Pb (with internal dilution)
Reagent Requirements: Keep between 10 - 30 °C
Sample Flow Rate: 100 - 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 1 year
Weight: 25 kg