CLF10 sc Free Chlorine Sensor, sc200 Controller and Stainless Steel Panel with pHD Differential Sensor, METRIC

OBSOLETE ITEM

CLF10 sc Free Chlorine Sensor, sc200 Controller and Stainless Steel Panel with pHD Differential Sensor, METRIC
Product #: 2981200

OBSOLETE ITEM

EPA-approved Free or Total Chlorine Analysis

The CLF10 sc and CLT10 sc chlorine analyzers are compliant with EPA Method 334.0, ensuring that you not only have precise process control, but reportable results as well.

No Reagents to Replace, No Waste Stream

Hach’s amperometric chlorine technology does not use reagents, eliminating the need for a sanitary drain and saving you the cost of monthly reagent replacement.

Chlorine Analysis Made Easy

The CLF10 sc and CLT10 sc are plug-and-play with Hach’s digital controllers, ensuring easy setup. Available with exclusive self-diagnostics, these instruments will alert the user to changes in the process or upcoming maintenance. Hach’s Cal Watch feature will provide a warning for pH or chlorine calibration changes, and a non-contacting flow sensor will notify users of insufficient sample flow.

What's included?

Hach sc200 Controller, Mounting Hardware, sc200 Basic User Manual, CL10sc Panel, 1 M digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.

Specifications

Accuracy: Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2: Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3: Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4: Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection: 5 pin, M12 connector
Cable Length: 1 m (between gateway and sc controller)
Calibration Method: 1-point or 2-point (zero and slope) calibration
Certifications: CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Connection Drain Line: 10 mm OD
Connections: 6 mm OD
Controller: Single Digital
Controller Compatibility: sc controller platform
Controller Options: Single Input
Description : CLF10sc Free Chlorine Sensor, SC200 Controller and Stainless Steel Panel with pHD Differential Sensor, METRIC
Dimensions : mm x 25 mm (7.68 in x 0.98 in)
  Sensor Only: 195
Fitting type: Metric
Fitting Type: Metric
Interferences: Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Length: Sensor: 195 mm
Lower Limit of Detection (LOD): 30 ppb (0.03 ppm)
Material: Corrosion-resistant materials, fully-submersible
(stainless steel, PEEK, PVC, silicon rubber and polycarbonate)
Measurement method: Reagentless, electrochemical, three-electrode amperometric system
Mounting: Panel or Wall Mounting
Operating Temperature Range: 0 - 45 °C
Options: SC200 Controller, Stainless Steel Panel, & pHD Differential Sensor
Parameter: Free Chlorine
pH Monitoring Required?: with pHD Differentinal
pH Range: 4 - 9
Power requirements (Voltage): 110 - 230 V AC
Pressure Range: 0.5 bar, no pressure impulses and/or vibrations
Range: 0 - 20 ppm Cl₂
Repeatability: 30 ppb or 3%, whichever is greater
Response Time: Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
  Total Chlorine: 100 s or less for 90% change (T90)
at a stable temperature and pH
Sample Flow Rate: 30 - 50 L/h (7.9 - 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature: 5 - 45°C (41 - 113°F)
Storage conditions: -20 °C to 60 °C dry
Warranty: 12 months
Weight: 12.79 kg Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?: Hach sc200 Controller, Mounting Hardware, sc200 Basic User Manual, CL10sc Panel, 1 M digital extension cable, Panel Manual, pHD Sensor Manual, Chlorine Sensor Manual.