Ambient Temperature:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog Outputs:
Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8
Automatic cleaning:
Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Calibration:
Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Certifications:
CE, ETL certified to UL and CSA safety standards, UKCA
Cycle Time:
No Dilution
Default: 10 min.
Continuous: 6 min.
Internal Dilution
Default: 15 min.
Continuous 9 min.
Demineralised water:
Internal dilution option: Rinse water & Dilution water
Digital outputs:
Profibus DP or Modbus RTU
Dimensions (H x W x D):
688 mm x 460 mm x 340 mm
Drain:
Atmospheric pressure, vented, min. Ø 32 mm
Exceedance Tracking:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Interferences:
Acidity, hardness does not interfere < 400 mg/L (as CaCO3), metal ions, e.g., aluminum (III) (Al3+) > 10 mg/L, cyanide (CN–), iron(III) (Fe3+) > 10 mg/L, nickel (II) (Ni2+) and silver (II) (Ag+). Large quantities of color and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere.
Lower Limit of Detection (LOD):
No Dilution
0.003 - 0.3 mg/L Cu: 0.003 mg/L
0.02 - 0.75 mg/L Cu: 0.02 mg/L
0.02 - 1.5 mg/L Cu: 0.02 mg/L
0.03 - 3 mg/L Cu: 0.03 mg/L
Internal Dilution
0.15 - 15 mg/L Cu: 0.15 mg/L
0.3 - 30 mg/L Cu: 0.3 mg/L
0.75 - 75 mg/L Cu: 0.75 mg/L
1.5 - 150 mg/L Cu: 1.5 mg/L
2.25 - 225 mg/L Cu: 2.25 mg/L
3 - 300 mg/L Cu: 3 mg/L
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method:
Bicinchoninate colorimetric method
Number of sample streams:
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Outputs:
Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
Parameter:
Copper
Power:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Precision:
Better than 2% full scale range for standard test solutions
Protection Class:
IP44
Range:
No Dilution
0.003 - 0.3 mg/L Cu
0.02 - 0.75 mg/L Cu
0.02 - 1.5 mg/L Cu
0.03 - 3 mg/L Cu
Internal Dilution
0.15 - 15 mg/L Cu
0.3 - 30 mg/L Cu
0.75 - 75 mg/L Cu
1.5 - 150 mg/L Cu
2.25 - 225 mg/L Cu
3 - 300 mg/L Cu
Reagent Requirements:
Keep between 10 - 30 °C
Relays:
5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Sample Flow Rate:
100 - 300 mL/min
Sample Pressure:
By external overflow vessel (open to atmospheric pressure)
Sample Quality:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature:
10 - 30 °C
Validation:
Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Warranty:
12 months
Weight:
Max. 35 kg
What's included?:
EZ1010sc Copper Analyser, Instruction Manual, 1 x Double Bit Door key, 1 x Mounting Brackets and 2 x empty 5L Reagent Containers with Fittings (for Colour & Acid Solution)