EZ6200 Total Arsenic Analyser

EZ6200 Total Arsenic Analyser
Product #: EZ6200.XXXXXXXX
Unit Price Contact Hach

What's included?

EZ6200 Arsenic Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x Reference Electrode (Ag/AgCl), 1 x Working Electrode - Gold, 1 x Counter Electrode, 2 x empty 2.5L Reagent Containers with Fittings (Buffer & HNO₃ Solution) and 1 x empty 5L Reagent Container with Fittings (H₂SO₄ Solution) and 2 x empty 1L Reagent Containers with Fittings (Reference Blank & Reference As Solution)

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog Outputs: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Automatic cleaning: Yes
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / ETL certified
Cooling water: Flow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
Cycle Time: 20 minutes (dilution + 5 min.)
Demineralised water: For rinsing / dilution
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 32 mm
Exceedance Tracking: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Iodide, organic matter, copper Cu(II) > 30 µg/L, iron Fe(III) >20 mg/L, various metals in mg/L levels may interfere. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD): ≤ 1 µg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method: Stripping voltammetry using gold electrode
Model: EZ6200
Number of sample streams: 1, 2, or 4
Parameter: Arsenic, total
Power: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Precision: Better than 5% full scale range for standard test solutions
Protection Class: Analyser cabinet: IP44 / Panel PC: IP65
Range: 1 - 20 µg/L As (optional ranges with internal dilution)
  1 - 20 µg/L As

Optional:
5 - 80 µg/L (with internal dilution)
10 - 200 µg/L (with internal dilution)
20 - 400 µg/L (with internal dilution)
Reagent Requirements: Keep between 10 - 30 °C
Sample Flow Rate: 100 - 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 12 months
Weight: 25 kg
What's included?: EZ6200 Arsenic Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x Reference Electrode (Ag/AgCl), 1 x Working Electrode - Gold, 1 x Counter Electrode, 2 x empty 2.5L Reagent Containers with Fittings (Buffer & HNO₃ Solution) and 1 x empty 5L Reagent Container with Fittings (H₂SO₄ Solution) and 2 x empty 1L Reagent Containers with Fittings (Reference Blank & Reference As Solution)