Ambient Temperature:
5 to 40 °C (41 to 104°F)
Air conditioning and heating options are available.
Communication:
Modbus RTU, Modbus TCP/IP & Profibus (when the Profibus option is selected, the digital output signals are sent through the Profibus converter with its specific communication protocol)
Except for Zone 1 certification then Modbus RTU, Modbus TCP/IP & Modbus TCP/IP Redundant is available
Cycle Time:
From 6.5 minutes, depending on range and application
Data Storage:
Previous 9999 analysis data on screen in the microcontroller memory and storage of data archive for the lifetime of the analyzer in the SD/MMC card.
Previous 99 fault data on screen in the microcontroller memory and storage of fault data archive for the lifetime of the analyzer in the SD/MMC card.
Display:
High contrast 40 character x 16 line backlit LCD with CFL backlight
Drain Pressure:
Typically ambient (for applications with high drain pressure, optional systems are available)
EExp / Hazardous Location:
Certification options are available to European Standards, (ATEX Zone 1, Zone 2), North American Standards (Class I Division 2) and IECEx Zone 1
Enclosure Rating:
IP44; optional IP54 with air purge
Expected Measuring Range:
Standard range < 5000 mg/L
Humidity:
5 - 85 % , non-condensing
Languages:
English
Lower Limit of Detection (LOD):
0.06 mg/L C
Measurement method:
Infrared measurement of CO2 after oxidation
Multi-Stream:
1, 2, 4 and 6 streams
Number of Channels:
1
Number of streams:
1
Outputs:
4 - 20 mA
Oxidation Method:
Patented Two-Stage Advanced Oxidation Process using Hyroxyl Radicals.
Parameter:
TOC, TIC, TC, VOC, after correlation COD, BOD
Particle Size:
Up to 2 mm, soft particulates
Permissible Chloride Range:
Up to 30%
Power Requirements (Hz):
60 Hz
Power requirements (Voltage):
115 V AC
Power supply:
TOC/TIC (115VAC)
Range:
0 - 10000 mg/L C
Range Selection:
Automatic or Manual Range Selection
Repeatability:
± 3 % of reading or ± 0.3 mg/L C, whichever is greater, with Automatic Range Selection
Response Time:
TOC 6.5 minutes
Sample Inlet Temperature:
2 - 60 °C
Sample Volume:
Up to 8.0 mL
Service Interval:
6 months service intervals
User Interface:
Microcontroller with membrane keyboard
Warranty:
1 year
12 months
Weight:
90 - 120 kg
Enclosure weight may change depending on system optional features.
What's included?:
Includes: B7000i Analyser, Tubing, Fuses, Ferrules, Drain, Acid & Base Dip Tubes, CO₂ Filter & B7000i User Manual