Communication Capabilities:
1x Ethernet, 1x USB-A
Compliance Certifications:
European directives: Low voltage 2006/95/EC; EMC 2004/108/EC
Electromagnetic compatibility standards: EN61326
Safety standard: IEC/UL/CSA 61010-1
Safety of Laser products: IEC/UL/CSA 60825-1
Dimensions (H x W x D):
942 mm x 540 mm x 537 mm
Display Type:
Digital TFT VGA (640 x 480) colour touch screen
Enclosure Rating:
IP20
Installation Requirement:
Power requirements: 100/240 VAC, 50/60 Hz
Purge gas: CO2 with purity >99.9% at 6 to 7 bar abs (87 to 102 psi)
Laser Class:
A Class 1 LASER is installed in this instrument.
Exposure to the beam of a Class 1 LASER will not result in eye injury.
Intrabeam viewing of Class 1 laser products which emit visible radiant energy may still produce dazzling visual effects, particularly in low ambient light.
Operating Conditions:
Package temperature: -2 - 30 °C (28 - 86 °F) Package pressure: 1.4 - 6.8 bar (20 - 73 psi) Ambient temperature: 0 - 40 °C (32 - 104 °F)
Operating system:
Windows® CE 4.2
Package Type:
Closure types: metallic, PET
Package height:
min. 90 mm (3.54 in.)
max. 340 mm (13.39 in.)
Package volume: min. 150 mL
Power Consumption:
250 VA maximum
Power Requirements (Hz):
50/60 Hz
Power requirements (Voltage):
100/240 VAC
Range:
O2: 1 ppb - 12 ppm
CO2: 0.75 - 5 v/v or 1.5 - 10 g/kg
Repeatability:
TPO: ±5 µg/L / ±10%,
whichever is greater
CO2: ±0.05 v/v or 0.10 g/kg ±2 %, whichever is greater, measured on equilibrated packages at 10 - 25 °C
Typical Analysis Time:
Approximately 4 minutes
Units:
O2 Concentration: ppb, ppm
CO2 Concentration: v/v, g/kg, g/L %W
Total Package Quantities: mg, mL, mg/L,mL/L, µg/L
Pressure: bar, mbar, psi
Temperature: °F, °C, K
Warranty:
12 months
Weight:
55 kg
What's included?:
Orbisphere 6110 Total Package Analyzer (sensors pre-installed) with Installation Kit, User Manual. See user manual on downloads tab for installation kit components.