Accuracy:
Accuracy = Trueness (ISO 57251)
2935A-A: The greater of ±1% of reading or ± 10 ppb, or ± 20 Pa
2952A-A: The greater of ±1% of reading or ± 2 ppb, or ± 5 Pa
2956A-A: The greater of ±1% of reading or ± 0.1 ppb¹, or ± 1 ppb², or ± 0.25 Pa
29552A-A: The greater of ±1% of reading or ± 2 ppb, or ± 5 Pa
¹ Accuracy is ± 0.1 ppb for 410, 510, 362x, 360x and 3655 instruments
² Accuracy is ± 1 ppb for 366x and 3650 instruments
Body Material:
Stainless Steel
Certifications:
CE, Ex II 1 G, Ex ia IIC T6
Flow Rate:
Recommended flow rate in flow chamber
2935A-A: 25 mL/min
2952A-A: 50 mL/min
2956A-A: 180 mL/min
29552A-A: 50 mL/min
Material:
O-Rings: FKM/FPM
Model:
GA2800 EX
Pressure Range:
100 bar
Pressure resistance:
40 bar minimum with default PPS collar (100 bar with stainless steel collar)
Protection Class:
IP66
Range:
Range at 25 °C
2935A-A: 10 ppb - 400 ppm or 20 Pa - 1,000 kPa
2952A-A: 1 ppb - 80 ppm or 5 Pa - 200 kPa
2956A-A: 0.1 ppb - 20 ppm or 0.25 Pa - 50 kPa
29552A-A: 2 ppb - 80 ppm or 5 Pa - 200 kPa
Response Time:
Response time at 25 °C for a 90% signal change
2935A-A: 2.5 min
2952A-A: 38 s
2956A-A: 7.2 s
29552A-A: 90 s
Sensor material:
Stainless Steel
Temperature Compensation:
-5 - 60 °C
Temperature resistance:
-5 - 100 °C
Warranty:
6 months
Weight:
0.3 kg
Wetted Materials:
Stainless Steel 1.4404 (AISI 316L), surface finish: N5, Ra < 0.4 µm
No O-rings are in contact with the sample
What's included?:
Oxygen Sensor (GA280E-SVS), incl. plastic screw-on storage cap for sensor head protection; protection cap without grille (33051-S0); protection cap with grille (33051-SP)