New: Polymetron 2582sc Dissolved Oxygen Analyzer, Claros-enabled, LAN + Profibus DP, 100-240 VAC, without power cord

Polymetron 2582sc Dissolved Oxygen Analyzer, Claros-enabled, LAN + Profibus DP, 100-240 VAC, without power cord
Product #: 2582.99.AAC801
Unit Price Contact Hach

Save Time on Design

A single design source and one product platform means you spend less time searching for design files or configuring components.

Easy Adoption

The familiar experience with a modern touchscreen, the ability to use your current Hach sensor, and the same footprint as the 9500 Polymetron, make installation and integration of the SC4500 Controller seamless.

The Connectivity Options You Need

The Controller provides local communication to SCADA or a PLC. From analog, to advanced digital protocols, to wi-fi, cellular or LAN, the SC4500 gives you the flexibility to adapt in a rapidly changing world.

Simplify Maintenance and Operation

Common menu guides reduce variability and provide step-by-step procedures for maintenance and calibration. Standard visual alerts across parameters notify operators when troubleshooting is required. Start-up and maintenance time are minimised with pre-mounted membrane cap and factory pre-conditioned sensors.

What's included?

Controller, mounting hardware, oxygen sensor, temperature sensor, cable, flow-chamber, basic user manual

Specifications

Analog Outputs: Five 0-20 mA or 4-20 mA analog outputs on analog output module
Cable Length: 10 m (33 ft)
Calibration Method: Zero: Electrically or with oxygen free water, Slope: in air or against a laboratory measurement
Communication: Analog:
Five 0 - 20 mA or 4 - 20 mA analog outputs on each analog output module
Up to two analog Input modules (0 - 20 mA or 4 - 20 mA). Each input module replaces a digital sensor input.

Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
  Modbus RS232/RS485, Profibus DPV1, Profinet optional
Communication Capabilities: Claros LAN + Profibus DP
Conduit Openings: ½" NPT Conduit
Connection Drain Line: 8 mm tubing
Connections: ¼" NPT thread (6 mm or ¼" tubing advised)
Display: 3.5-inch TFT color display with capacitive touchpad
Electrical Certifications: CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, CMIM, Morocco
Enclosure Rating: UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X
Metal enclosure with a corrosion-resistant finish
Lower Limit of Detection (LOD): < 1 ppb
Maintenance Interval: Membrane Lifetime: 6 months depending on sample
Material: Electrodes: Gold cathode, Silver anode; Flow-through cell: Stainless Steel 316L; Membrane: PFA, Sensor Body: Noryl
Material Enclosures: Polycarbonate Aluminium (powder coated) Stainless Steel
Measuring range: 0 to 2000 ppb (0-2 ppm)
Operating Temperature Range: -20 - 60 °C (-4 - 140 °F) (8 W (AC)/9 W (DC) sensor load)
-20 - 45 °C (-4 - 113 °F) (28 W (AC)/20 W (DC) sensor load)
Linear derating between 45 - 60 °C (-1.33 W/°C)
Power Requirements (Hz): 50/60 Hz
Power requirements (Voltage): 100 - 240 VAC
Power supply: Without power cord
Range: 0 - 2000 ppb (0 - 2 ppm)
Relay: Operational Mode: Primary or secondary measurement or timer/scheduler
Relay Functions: Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
Relays: Two relays (SPDT);

Wire gauge: 0.75 - 1.5 mm² (18 - 16 AWG)

AC controller
Maximum switching voltage: 100 - 240 VAC
Maximum switching current: 5 A Resistive/1 A Pilot Duty
Maximum switching power: 1200 VA Resistive/360 VA Pilot Duty

DC controller
Maximum switching voltage: 30 VAC or 42 VDC
Maximum switching current: 4 A Resistive/1 A Pilot Duty
Maximum switching power: 125 W Resistive/28 W Pilot Duty
Repeatability: ±0.5 ppb or ±5%, whichever is greater
Reproducibility: ±0.5 ppb or ±2% whichever is greater
Response Time: For step change 1 - 40 ppb: < 30 s
Sample Flow Rate: 66 - 166
Security Levels: 2
Software available: Cloud (Claros-enabled)
Storage conditions: -20 - 70 °C (-4 - 158 °F)
Temperature Compensation: Automatic in the range of 0 - 45 °C (32 - 113 °F)
Units: mg/L, ppm, µg/L, ppb, mbar, hPa, inch Hg, mmHg
Warranty: 12 months
Weight: 3.2 kg
What's included?: Controller, mounting hardware, oxygen sensor, temperature sensor, cable, flow-chamber, basic user manual