Ambient Temperature:
-20 - 60 °C
Analysis time:
Approx. 4 minutes
Dimensions :
440 mm x 380 mm x 700 mm (H x W x D)
Gas consumption:
200 mL/min*
*Sample preparation for NPOC determination requires additional 300-350 mL/min.
Gas specifications:
Oxygen or synthetic air: minimum 99.998% (4.8) at 3 - 10 bar
Lower Limit of Detection (LOD):
TC, TIC, NPOC: 50 µg/L
TNb: 20 µg/L
Measurement method:
TOC: NDIR (non-dispersive Infrared Detection)
TN: Chemiluminescence
Model:
QP1680-TOC/TNb
Norms and Standards:
TOC / NPOC: ASTM D7573, EN 1484, EPA 415.1, EPA 9060A, ISO 8245, SM 5310B, NEN-ISO 20236
TNb: ASTM D8083, EN 12260, ISO 11905-2, NEN-ISO 20236
Number of sample positions:
65
Operating Conditions:
20 - 30 °C; 20 - 80% relative humidity (non-condensing)
Oxidation Method:
Catalytic combustion at 720 °C
Parameter:
TOC, TNb
Pollution Degree:
2
Power:
Analyser: 750 W max.
PC: 90 W max.
Monitor: 100 W max.
Power supply:
Analyser: 100 - 240 VAC, 50/60 Hz, 16 A, with protective grounding
PC: 100 - 240 VAC, 50/60 Hz, 1.6 A, with protective grounding
Monitor: 100 - 240 VAC, 50/60 Hz, 1.6 A, with protective grounding
Range:
TC, TIC, NPOC, TNb: 0 - 30000 mg/L
Repeatability:
Up to 10 mg/L TC, TIC, NPOC, TN: < 5%
> 10 mg/L TC, TIC, NPOC, TN: < 2%
Sample Volume:
10 - 1000 µL
Temperature:
Furnace temperature max. 1050 °C (depending on configuration)
Warranty:
12 months
Weight:
Approx. 32 kg
What's included?:
TOC/TNb Instrument
Sample tray with 65 positions for 40 mL vials
Glass sample vials 40 mL (100 pcs)
USB 2.0 cable A-B
Set of tubing PTFE blue-stripe 1/8" OD length 2 m including nut & ferrule
Reagent container with screwcap for UPW, Acid and Waste
Power cord
USB storage device with software, test results and Full User Manual
Printed User Manual