Cable Length:
1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration Method:
For Formazin and Stablcal:
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range
Custom calibration for up to 6-point calibrations
Certifications:
CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Australian ACMA Marking
Controller:
SC4500
Dimensions (H x W x D):
685.8 mm x 584.2 mm x 292.1 mm
Fitting type:
Sample quick connector: ¼-in. for ¼-in. tubing
Light source:
Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Material:
ASA Luran S 777K / RAL7000, TPE RESIN Elastocon STK40,
Thermoplastic Elastomer TPS-SEBS
Mounting:
Wall or rack mounting
Mounting Configurations:
Wall Mount
Operating Humidity:
Relative humidity: 5 - 95% at different temperatures, non-condensing
Power requirements (Voltage):
100-240 VAC ±10%, 50/60 Hz; 1 A
Process connection:
Sample drain: 1/4" OD tube or hose barb
Sample Flow Rate:
100 - 1000 mL/min, optimal flow rate: 200 - 500 mL/min
Sample Pressure:
6 bar (87 psi) maximum, compared to air at sample temperature range from 2 - 40 °C (35.6 - 104 °F)
Sample Temperature:
2 to 60 °C (35 to 140 °F)
Sensor to Controller Distance (Maximal):
1000 m (3280 ft), per RS485
Sensor Type:
Digital output
Storage conditions:
-40 to 60 °C (-40 to 140 °F)
Stray Light:
<10 mNTU
Units:
NTU, FNU, TE/F, FTU, EBC
Verification:
Liquid: Stablcal, Formazin (0.1 to 40 NTU)
Dry: Glass Rod at <0.1 NTU
Weight:
13.8 kg / 30.5 lbs
What's included?:
TU5400sc Turbidimeter (LXV445.53.B3212)
SC4500 Controller, Claros-enabled, 5x mA Output, 2 digital Sensors, 100-240 VAC, US plug (LXV525.99EA1551)
Flow Sensor for TU5300sc and TU5400sc Laser Turbidimeter (LQV160.99.00002)