Accuracy:
Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic Wiper:
Yes
Body Material:
Stainless Steel
Cable Length:
10 m (optional extension cables available)
Calibration Method:
Turbidity: Formazin or STABLCAL Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Certifications:
CE
Controller:
Sensor only
Diameter Sensor:
60 mm
Flow:
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length:
200 mm
Maintenance Interval:
1 h/month
Material:
Optics carrier and sleeve: PVC black
Wiper arm: stainless steel 1.4581
Wiper rubber: silicone rubber (standard)
Optional: FKM/FPM (LZX578)
Wiper shaft: stainless steel 1.4571
Max. cable length Sensor-to-Controller:
100 m
Measuring Principle:
Infrared Duo scattered light technique for color-independent turbidity measurement Turbidity in accordance with DIN EN ISO 7027 / TS equivalent DIN 38414
Model:
Solitax ts-line sc
Mounting:
Immersion
Operating Temperature Range:
0 - 40 °C
Parameter:
Suspended Solids, Turbidity
Pressure Range:
Stainless steel immersion sensor: <6 bar or <60 m
Range:
TSS content: 0.001 mg/L - 50 g/L
Turbidity: 0.001 - 4000 NTU
Repeatability:
TSS content: < 3 %
Turbidity: < 1 %
Response Time T90:
1 - 300 s adjustable
Special notes:
installation note: distance sensorwall > 50 cm (Turbidity);
distance sensorwall>10 cm (TSS),
Warranty:
12 months
Weight:
1.38 kg
What's included?:
Turbidity/Suspended Solids probe, user manual