Accuracy:
±2% of reading plus 0.01 FNU/NTU from 0 - 1000 FNU/NTU
Air purge:
Dry nitrogen or instrument grade air (ANSI MC 11.1, 1975)
0.05 L/s at 69 kPa (10 psig); 138 kPa (20 psig) max
Hose barb connection for 1/8-inch tubing
Certifications:
CE, KC, RCM
Communication:
USB
Compliance:
ISO
Data logging:
2000 total logs, includes reading log, verification log and calibration log
Dimensions (H x W x D):
153 mm x 395 mm x 305 mm
Display:
17.8 mm colour touch screen
Interface:
2 USB-A ports for USB flash drive, external printer, keyboard and barcode scanner
Light source:
Light-emitting diode (LED) at 860 ± 30 nm
Measurement method:
Nephelometric
Model:
TL2310 ISO
Operating Temperature Range:
0 - 40 °C
Power requirements (Amps):
3.4 A
Power Requirements (Hz):
50/60 Hz
Power requirements (Voltage):
100 - 240 VAC
Range:
0 - 1000 NTU
NTU/FNU: 0 - 1000
Reading modes:
Single, continuous, Rapidly Settling Turbidity, signal averaging on or off
Regulatory:
Meets ISO 7027, DIN EN 27027, DIN 38404 and NFT 9033
Repeatability:
±1% of reading or 0.01 FNU/NTU, whichever is greater (under reference conditions)
Response Time:
Signal averaging off: 6.8 seconds / Signal averaging on: 14 seconds (when 10 measurements are used to calculate the average)
Sample cell compatibility:
Round cells 95 x 25 mm (3.74 x 1 in.) borosilicate glass with rubber-lined screw caps
Sample Requirements:
25 mm sample cell: 20 mL minimum
0 to 70 °C (32 to 158 °F)
Source Lamp:
LED
Stabilisation time:
Immediately
Storage conditions:
-20 to 60 °C
Units:
FNU and NTU
Warranty:
12 months
Weight:
3.0 kg
What's included?:
TL2310 Turbidimeter, silicone oil, oiling cloth, 1-inch sample cells (30 mL) with caps (6x), Gelex secondary turbidity standardization kit, Stablcal calibration kit, power supply, power cord, dust cover