TSS Ex sc, Suspended Solids Inline Sensor without a Wiper

TSS Ex sc, Suspended Solids Inline Sensor without a Wiper
Product #: LXV328.99.30002
Unit Price Contact Hach
Ships within 3-5 weeks

Unique Multi-Beam Pulsed Infrared Light System

TSS sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity from 0.001 to 4,000 FNU and suspended solids from 0.001 to 500 g/l.

Automatic Compensation for Air Bubbles and Temperature Swings for Accuracy

Special software enables the system to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.

Standard Compliant and Precise - Ease of Calibration

The turbidity measurement complies with the standard DIN EN 27027 (ISO 7027). Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.

Withstands Difficult Conditions and Harsh Environments

Probes are made of highly polished stainless steel with a scratch resistant sapphire window. They are design to withstand harsh environments and keep particles from sticking to the surface.

Suitable for Use in Potentially Explosive Atmospheres

TSS EX1 sc probes have been specially designed for measurement in hazardous locations with Class I Div 2 requirements.

What's included?

Turbidity & Suspended Solids sensor, user manual

Specifications

Accuracy: Turbidity up to 1000 NTU:
< 5% of the measured value ±0.01 NTU
Ambient Temperature: 0 - 50 °C
Application: Hazardous locations
Automatic Wiper: No
Cable Length: 10 m
  10 m (optional extension cables available)
Calibration: Turbidity (TRB): Factory calibrated
Solids (TS): To be calibrated by customer on site
Zero point: Permanently calibrated in the factory
Calibration Method: Turbidity Formazin or StablCal® Standard (at 800 NTU). Requires a calibration kit.
Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure.
Controller Compatibility: SC200, SC1000, SC4500. All controllers sold separately
Diameter: 48.5 mm
Flow: Max. 3 m/s (the presence of air bubbles affects the measurement)
Includes: Turbidity & Suspended Solids sensor, user manual
Installation style: Ball Valve Insertion
Length: 488 mm
Maintenance Interval: 1 h/month
Material: Optics Carrier and Sleeve: stainless steel 1.4460 / stainless steel 1.4404
Max. cable length Sensor-to-Controller: 100 m
Max Temperature: 50 °C
Measurement method: Combined multiple beam alternating light method with infrared diode system and beam focusing
Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm
Solids (TS): 90º and 120º scattered light measurement, wavelength = 860 nm
Measuring Principle: Turbidity: 2-channel 90° scattered light measurement in accordance with DIN/EN 27027/ISO7027
TSS: Modified absorption measurement
Eight-channel multiple-angle measurement, wavelength = 860 nm
Model: TSS EX1 sc
Mounting Configurations: Insertion
Operating Temperature Range: -10 - 50 °C
Parameter: Turbidity, Suspended Solids
Pressure Range: 10 bar or 100 m
Range: Turbidity (TRB): 0.001 - 9999 NTU
Solids (TSS): 0.001 - 500 g/L
Repeatability: TSS content: < 4 %
Turbidity: < 3 %
Response Time: 1 s < T90 < 300 s (adjustable)
Response Time T90: 1 - 300 s adjustable
Special Feature: Simple Inline installation
Special notes: Installation note:
Distance sensor–wall
> 10 cm (TSS)
Special Notes: Installation note:
Distance sensor–wall
> 50 cm (Turbidity)
Warranty: 12 months
Weight: approx. 2.7 kg
Weight Sensor: 2.7 kg
What's included?: Turbidity & Suspended Solids sensor, user manual