Accuracy:
Using standard solutions: 3% + 0.02 mg/L ; > 1.0 mg/L: 5% + 0.02 mg/L
Cable Length:
2 m fixed data cable at analyzer
2 m fixed power cable at analyzer
Dimensions (H x W x D):
720 mm x 540 mm x 390 mm
Enclosure Rating:
IP55
Filtersystem included:
No, use e.g. Filtrax
Flow:
1 - 20 L/h Sample (free of suspended solids)
Fuse Protection:
Via SC1000 controller or power box. Maximum of 2 analysers per SC1000 or power box
Indoor/Outdoor:
Outdoor
Lower Limit of Detection (LOD):
0.02 mg/L NH4-N
0.02 mg/L NH4–N
Material:
ASA/PC UV-resistant
Measurement method:
GSE (Gas Sensitive Electrode) with screw-on membrane cap
Measuring Interval:
5 - 120 min, adjustable per 5 min.
Measuring Principle:
NH4 is being converted to NH3 gas, wich is being measured at the GSE.
Number of Channels:
2
Operating Humidity:
95% relative humidity, non-condensing.
Operating Temperature Range:
-20 - 45 °C
Outputs:
Current output, relays and bus communication via SC controller
Parameter:
Ammonia
Permissible Chloride Range:
Max. Cl- concentration: 1000 mg/L
pH Range:
5 - 9 pH
Power Requirements (Hz):
50 Hz
Power requirements (Voltage):
115 - 230 V AC, powered by Power Box or SC Controller
Pressure Range:
-30 - 50 mbar With continuous sample preparation; at overflow vessel
Range:
0.02 - 5.0 mg/L NH₄-N
Reproducibility:
2% + 0.02 mg/L
Response Time:
0.02 - 2.0 mg/L: 15 min;
2.0 - 5.0 mg/L: 5 min
Sample Level:
Sample level in basin and filtration system must be below analyzer.
Sample Quality:
Ultra filtrated or comparable.
Sample Temperature:
4 - 40 °C (39 - 104 °F)
Sampling Option:
two channel continuous sample from filtration unit, e.g. FILTRAX
Special Feature:
Lockable housing; automatic cleaning and calibration; extensive self-diagnostics
Storage conditions:
4 - 55 °C; 95% relative humidity, non-condensing
Weight:
Without sample preparation system and without chemicals: 31 kg
What's included?:
0.02 - 5 mg/L NH₄-N, 115 VAC Amtax sc Ammonia Analyzer, 3 month supply of all necessary consumables at 5 minute cycle time, user manuals.