HQ40d Portable Meter Kit with PHC301 pH Electrode and CDC401 Conductivity Cell

OBSOLETE ITEM

HQ40d Portable Meter Kit with PHC301 pH Electrode and CDC401 Conductivity Cell
Product #: HQ40D53151201

OBSOLETE ITEM

Portable meter measures critical water quality parameters - without the need for multiple instruments

Two input channels for simultaneous measurement of pH, Conductivity, Dissolved Oxygen, BOD, ORP, Ammonia, Ammonium, Fluoride, Chloride, Sodium, and temperature - any IntelliCAL™ smart probe

Intuitive user interface for simple operation and accurate results

Guided calibration and check standard routines reduce calibration errors. Stabilization alerts and visual measurement lock ensure that you can trust in the accuracy of the results.

Trust your measurements - IntelliCAL™ smart probes store all calibrations in the probe

Calibration history allows quick and easy change out of probes without re-calibrating. The HQd smart system records serial numbers, current calibration data, user ID, sample ID, time, and date automatically in the data log for complete GLP traceability.

Designed for demanding conditions

Rugged, waterproof (IP67) meter, protective glove, and rugged field case provide worry-free, reliable operation in lab or field environments.

Convenient kit includes everything you need to start testing

Meter kit includes CDC401 conductivity cell with 1m cable, PHC301 refillable pH electrode, field case, 4 AA batteries, protective meter glove, power adapter, USB/DC power adapter for data transfer, quick start guide, and user manuals

What's included?

Includes: meter, conductivity probe with 1 meter cable, refillable pH electrode with 1 meter cable, 4 AA batteries, field kit (protective meter glove with straps, sample containers (5), meter caps), power cord, power supply, USB/DC adapter for data transfer, quick start guide, and user manuals

Specifications

AC Power Adapter and USB/DC Adapter: Included
Automatic Buffer Recognition: Yes
Backlight:
Barometric Pressure Measurement: Automatic compensation of DO when using an LDO or LBOD probe
Calibration: Demal (1D/ 0.1D/ 0.01D);
Molar (0.1M/ 0.01M/0.001M);
NaCl (0.05%; 25µS/cm; 1000µS/cm; 18mS/cm);
Standard sea water;
User defined
Calibration Intervals/Alerts/Reminder: Off, selectable from 2 hours to 7 days
Compliance Certifications: CE.WEEE
Conductivity measurement: Yes
Conductivity measurement: Temperature correction: None; Linear; NaCl Non-Linear Natural Water.
Conductivity Measurement Range: 0.01 µS/cm - 200.0 mS/cm
Conductivity resolution: 0.01 µS/cm - 0.1 mS/cm upon selected measuring range
Contents: No
Custom Calibration Standards: Yes
Data Export: Download via USB connection to PC or flash memory device. Automatically transfer entire data log or as readings are taken.
Data Memory: 500 records/FIFO
Data Storage: Automatic, GLP ISO compliant reading data stored with calibration details.
Digital (intelligent) electrode inputs: 2 channels
Display: Detailed mode/Large mode
Display Lock Function: Continuous / Auto-stabilization ("press to read") / At Interval
Display Type: 240 x 160 pixel LCD with backlight illumination
DO Measurement Range: 0.1 - 20.0 mg/L (ppm) 1 - 200% saturation
DO Resolution: 0.1
DO sensor calibration: * 100% (water-saturated air (100%) calibration
* 100% with 0 (water-saturated air (100%) calibration with 0 point
* mg/L (calibration with a specified dissolved oxygen concentration (mg/L) solution)
* mg/L with 0 (calibration with a specified dissolved oxygen concentration (mg/L) solution with 0 point)
* Factory (calibration with the default LDO calibration)
Electrode stand:
GLP Features: Date; Time; Sample ID; Operator ID
Inputs: 2
Instrument: Portable
IP Rating: IP67
ISE Direct Measurement Range: Yes
ISE Electrode Calibration: 2 - 5 Points
Kit?: Yes
Languages: English, French, German, Italian, Spanish, Danish, Dutch, Polish, Portuguese, Turkish, Swedish, Czech, Russian
Measurement method: Probe specific programmed method settings
Model: HQ40D – Multi/2 Channels
mV Measurement at Stable Reading: Yes
mV Measurement Range: -1500 - 1500 mV
mV Resolution: 0.1 mV
Needed Cable Length: 1
Operating Error Messages: Clear text error messages displayed
Operating Humidity: 90 % relative humidity (non-condensing)
Operating Interface: Soft Touch Keypad
Operating Temperature: 0 - 60 °C (32 - 140 °F)
ORP Electrode Calibration: Predefined ORP standards (including Zobell's solution)
Parameter: pH/Oxydo Reduction Potential (ORP)
Conductivity/Total Dissolved Solid (TDS)/Salinity/Resistivity
Dissolved Oxygen (DO)
Biochemical Oxygen Demand (BOD)
Ion Selective Electrode (ISE): Ammonia, Ammonium, Chloride, Fluoride, Nitrate, Sodium
Parameters: pH

Conductivity
NA
PC Data Transfer Software: HQD Series Meter Data Transfer Utility
pH Buffer Sets: IUPAC standards
(DIN 19266) or Technical buffer (DIN 19267) or 4-7-10 series or user defined
pH Electrode calibration: 1 - 3 Calibration points
Calibration summary data logged and displayed
pH Measurement Range: 0 - 14 pH
pH Resolution: Selectable:
0.001/0.01/0.1 pH
Printer: Yes, Optional
Probes included: PHC30101, CDC40101
Probe Type: Intellical Standard Laboratory or Rugged Field
Salinity Measurement Range: 0 - 42 (ppt) (‰)
Salinity Resolution: 0.01 (ppt) (‰)
Sensor A: PHC30101 (1 m)
Sensor B: CDC40101 (1 m)
Sensor C: NA
Sensors: Sensor A: PHC30101
Sensor B: CDC40101
Sensor C:
Simultaneous measurements: Yes, 2 channels
TDS Measurement Range: 0.00 mg/L - 50.0 g/L NaCl
TDS Resolution: 0.01 mg/L up - 0.1 g/L upon measuring range
Temperature Compensation: Automatic Temperature compensation for pH
  Automatic Temperature compensation for pH
Temperature Measurement: °C or °F
Temperature Resolution: 0.1
Type of measurement: Field Measurements
Warranty: 36 months
Weight: 335 g (0.75 lb) without batteries; 430 g (0.95 lb) with
What's included?: Includes: meter, conductivity probe with 1 meter cable, refillable pH electrode with 1 meter cable, 4 AA batteries, field kit (protective meter glove with straps, sample containers (5), meter caps), power cord, power supply, USB/DC adapter for data transfer, quick start guide, and user manuals