Accuracy:
With Standard Solution:
2% of the measured value + 0.015 mg/L
Altitude:
2000 m maximum
Cable Length:
Standard: 2 m cable fixed from the edge of the enclosure
Extension cables are available as an option (data cable up to 50 m, power cable up to 5 m).
Certifications:
CE, UKCA, CMIM, FCC, ISED, certified to UL and CSA safety standards by TÜV
Controller Compatibility:
Digital SC Controller. Controllers are sold separately.
Dimensions (H x W x D):
991 mm x 575 mm x 425 mm
Enclosure Rating:
Rating: IP55, NEMA 3R UL50E
Material: PUR 66
Filter Configuration:
Use e.g. Filtrax
Indoor/Outdoor:
Indoor or outdoor use
Lower Limit of Detection (LOD):
0.015 mg/L PO₄-P
Measurement method:
Photometric (Ortho-phosphate ions react with vanadate-molybdate reagent and form a yellow dye.)
Measuring Interval:
10, 15, 20 or 30 minutes (user adjustable)
Number of Channels:
2
Operating Temperature Range:
-20 - 45 °C; 95% relative humidity, non-condensing
Options:
Standard configuration
Outputs:
Relay, analog outputs, network interface through SC Controller
Parameter:
Orthophospate
Permissible Chloride Range:
≤ 1000 mg/L Cl-
pH Range:
5 - 9 pH
Pollution Degree:
2
Power requirements (Voltage):
115 V AC
Power supply:
Mains power supplied by SC Controller or LQV155 power box.
Analyser and heated drain tubing: 115 VAC
Protection Class:
Class I
Range:
0.015 - 5.0 mg/L PO₄-P
Repeatability:
With Standard Solution:
0.7% of the measured value + 0.005 mg/L
Response Time T90:
90% per measurement cycle for PO₄–P > 0.2 mg/L
80% per measurement cycle for PO₄–P ≤ 0.2 mg/L
Sample Flow Rate:
0.5 - 20.0 L/h
Sample Pressure:
2.5 bar (0.25 mPa) maximum
Sample Quality:
Ultra filtrated or comparable
Sample Temperature:
4 - 40 °C
Special Remarks:
The liquid level in the basin must be below the bottom of the analyser.
Storage conditions:
-20 - 60 °C; 95% relative humidity, non-condensing
Warranty:
12 months
Weight:
Approximately 45 kg without chemicals
What's included?:
NP6000sc Phosphate Analyser, Mounting Hardware, User Manual