Calibration Method:
Slope: using a laboratory reference value (e.g. LCW025)
Communication (optional):
Analog:
Five 0-20 mA or 4-20 mA analog outputs on each analog output module
Up to two analog Input modules (0-20 mA or 4-20 mA). Each input module replaces a digital sensor input.
Digital:
Profibus DPV1 module
Modbus TCP
Profinet IO module
Ethernet IP module
Communication Capabilities:
Claros Profinet
Compliance Certifications:
CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C (Morocco)
Connection Drain Line:
6 x 8 mm (tubing must not exceed 4 feet and must drain straight down)
Connections:
4 x 6 mm stainless steel tubing
Electrical Certifications:
EMC
CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Safety
CAN/CSA C22.2 No. 61010-1
cETLus safety mark for:
- General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1
Enclosure Rating:
UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X
Metal enclosure with a corrosion-resistant finish
Flow:
166 - 250 mL/min (10 - 15 L/h) recommended
Lower Limit of Detection (LOD):
Drift is negligible; 1 ppb
Maintenance Interval:
Monthly: Calibration and reagent refill
Manual Languages:
Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
Material:
Counter-Electrode: Stainless steel; Working Electrode: Platinum
Measuring range:
0 - 500 ppb hydrazine; programmable
Number of Channels:
1
Operating Temperature Range:
5 - 45 °C at 0 - 95% relative humidity (non-condensing)
Power requirements (Voltage):
110 - 240 V AC
Power supply:
Without power cord
Pressure Range:
0.5 - 6 bar (7.2 - 87 psi) or 12 L/h
Range:
0 - 500 ppb hydrazine; programmable
Range 2:
0 - 100 ppb carbohydrazide (also known as ELIMIN-OX); programmable
ELIMIN-OX is a registered trademark of Nalco Chemical Co., Naperville, IL.
Reagent:
Sample conditioning with diisopropylamine, ammonia, or diethylamine
Reference Type:
Ag/AgCl electrode with electrolyte KCl 0.1 M
Relative Humidity:
10 - 90% relative humidity (non-condensing)
Relay: Operational Mode:
Primary or secondary measurement or timer/scheduler
Relay Functions:
Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
Relays:
Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Repeatability:
± 2% or 1 ppb (whichever is greater)
Response Time:
< 60 s
Sample Requirements:
Sample needs to be free of undissolved matter.
Sample Temperature:
5 - 45 °C
Security Levels:
2
Software available:
Cloud (Claros-enabled)
Storage conditions:
-20 - 70 °C, 0 - 95% relative humidity, non-condensing
Weight:
14.6 kg
What's included?:
Controller installed in Panel, Working Electrode, Reference Electrode, Counter Electrode, Basic User Manual