Accuracy:
Turbidity: Defined according to ISO/WD 13530. Suspended Solids Less than of reading or ±0.001 NTU, whichever is greater
Accuracy 2:
Defined according to ISO/WD 13530. Turbidity Suspended Solids Less than 5 % of reading (depends on homogeneity of municipal activated sludge)
Automatic Wiper:
Yes
Cable Length:
10 m (33 ft.) standard. Optional extension cables available in 7.6 m (25 ft.), 15.2 m (50 ft.), 30.5 m (100 ft.). Maximum total length: (328 ft.).
Calibration:
Turbidity Suspended Solids:Formazin or StablCal® Standard
Based on gravimetric TSS analysis with a correction factor procedure
Calibration Method:
Turbidity Suspended Solids:Formazin or StablCal® Standard
Based on gravimetric TSS analysis with a correction factor procedure
Certifications:
CE certified to EN 61326-1, EN 61326/A1, EN 61326/A2, EN 61010-1
Construction - Sensor Body (RD 240/260):
Wiper
Sensor body: Silicon
Stainless steel or PVC
Controller:
sc200
Detection Limit:
Turbidity
Suspended Solids 0.001 mg/L
Diameter Sensor:
(diameter x lengImmersion sensor: 60 mm x (2.4 x)
Flow:
Flow Velocity 3 m/s (9.8 ft./s) maximum
Length:
200 mm
Lower Limit of Detection (LOD):
Turbidity
Suspended Solids. 0.001 NTU
Material:
Optics Carrier and Sleeve: stainless steel 1.4571 or black PVC
Wiper Arm: stainless steel 1.4581; Wiper Blade: silicone (standard) Optional: Viton (LZX578); Wiper Shaft: stainless steel 1.4571
Threaded cable fitting: stainless steel 1.4305 or white PVC
Stainless steel can be susceptible to corrosion at chloride concentrations ≥ 500 mg/l.
Max. cable length Sensor-to-Controller:
100 m
Measuring Principle:
Dual beam infrared/scattered light photometer to measure turbidity. A backscatter photoreceptor to measure suspended solids
Measuring Range Turbidity:
Suspended Solids 0.001 - 4000 NTU
Model:
Solitax hs-line sc
Mounting:
Fixed to the rim of the tank
Through sidewall of a pipeline using a ball valve; minimum pipe size (4 in.) in carbon or stainless steel
Operating Temperature Range:
0 - 40 °C
Parameter:
Hight Range Suspended Solids,
Turbidity
Pressure Range:
Sample Pressure 6 bar (87 psi) maximum
Range:
Turbidity
Suspended Solids 0.001 mg/L - 50 g/L
Repeatability:
<3 %
Repeatability Turbidity:
<1 % of reading
Response Time:
Initial response in 1 second
Sample Temperature:
0 - 40°C (32 - 104°F)
Signal Average Time:
Initial response in 1 second
Turbidity:
Units of Measure Turbidity
Suspended Solids user selectable—NTU, FNU, or TE/F
Units:
Units of Measure
Turbidity
Suspended Solids: User selectable—g/L, mg/L, ppm, or % solids
Weight:
1.38 kg
Weight Sensor:
Immersion, stainless steel: 1.38 kg (3.0 lb.)
What's included?:
Turbidity probe, controller, user manual