APA6000 Alkalinity Analyzer

OBSOLETE ITEM

APA6000 Alkalinity Analyzer
Product #: 5100010

OBSOLETE ITEM

Accurate, continuous analysis

The APA6000 provides an accurate measurement of alkalinity for your process.

Completely automated

The analyzer autocalibrates, set-and-forget operation frees time for other tasks.

Very low maintenance

The analyzer is self-cleaning, self-calibrating, and selfpriming. Reduced sample volume minimizes reagent use and allows instrument to run up to 30 days in clean water applications before maintenance and reagent replenishment.

Advanced process control

Two 4 - 20 mA outputs and two alarm relays allow quick configuration into standard PID control loop. Can be expanded up to 14 4-20 mA outputs and 14 alarm relays.

Constant protection

Fast and accurate response minimizes process excursions.

What's included?

APA6000 Alkalinity analyzer, installation kit, one month's supply of reagents, a maintenance kit, Hach user manual.

Specifications

Accuracy: Better than ± 5 % of reading or ± 1 mg/L whichever is greater
Analog Outputs: Two 4-20 mA outputs suitable for recorders or PID control. Output span programmable over any portion of the 1 to 500 mg/L range (130 Vac isolation from earth ground)
Calibration Interval: 60 minutes
Certifications: NRTL certified to UL and CSA standards and CE approved
Cycle Time: 8 minutes
Data Communications Distance: Maximum node-to-node distance is 400 m (1312'). Maximum total wire length is 500 m (1640')
Depth: (21")
Dimensions (H x W x D): 522 mm x 627 mm x 526 mm
Drain: Gravity, air break or vent recommended
Drain Fitting: 1/4" OD NPT barbed hose fitting
Enclosure Rating: NEMA 4X (indoor)/IEC 529 (IP66) with provision for air purge. Reagent enclosure is drip-proof.
Height : (21")
Inlet: Sample filter: 3/4" NPT male or female
Mounting: Wall, bench and panel
Parameter: Alkalinity
Power Requirements (Hz): 50/60 Hz
Power requirements (Voltage): 90 - 240 V AC
Range: 1 - 1000 mg/L total alkalinity as CaCO3
5 - 1000 mg/L CaCO3 as phenolphthalein alkalinity
Relay output: Two SPDT relays with contacts rated for 5A resistive load at 230 Vac. Additional relays available through optional Signal Output Module.
Repeatability: Better than 3 % of reading or ± 0.6 mg/L, whichever is greater
Response Time: Less than 10 minutes for 90% response to setup change at instrument sample fitting
Sample Flow Rate: 20 - 1000 mL/min
Sample Interval: Optionally capable of analyzing up to three sample streams
Sample Pressure: 0.03 - 6.9 bar
Sample Temperature: 5 - 50 °C (41 - 122 °F)
Weight: 25.5 kg
What's included?: APA6000 Alkalinity analyzer, installation kit, one month's supply of reagents, a maintenance kit, Hach user manual.
Width: (25")